Changing the high/low limit value
The following is an example to change the high limit value from +105% to +103%.
To change other high/low limit value setting, select the following functions.
Low limit value
J4: L ALM SET
High-high limit value
K2: HH ALM SET
Low-low limit value
K4: LL ALM SET
Switch operation
Display example
Description
J2: H ALM VAL
+105.0%
Select J2: H ALM VAL from the setting item selection menu.
The currently set high limit alarm value appears.
J2: H ALM VAL
+105.0%
The cursor appears.
Press
further to move the cursor to the digit to change.
J2: H ALM VAL
+103.0%
Change the number of the digit.
To change other digit, press
under this condition to move
the cursor to that digit and change the number there.
J2: H ALM VAL
+103.0%
The display blinks and the set value is displayed for
conformation
Press
MENU/ENT
again to write the data. Then the cursor
disappears and the display returns to the condition that the
setting value is displayed.
J2: H ALM VAL
Returns to the menu of setting item selection.
Note: If a value of 10% or less, or a value of 110% or more is tried to set, *L. Over SPEC. or *H. OVER SPEC.
appears and the value returns to the previously set value.
Try to set a new value again.