3 initial settings when shipped from the factory – Toshiba Tohsiba Electromagnetic Flowmeter Converter L5232 User Manual
Page 113
6F8A0917
- 112 -
8.3 Initial Settings When Shipped from the Factory
Initial settings used when shipped from the factory unless otherwise specified are shown below.
No.
Parameter name
Initial setting values
A1 Exciting
current
Adjusted value when shipped from the factory
(described on the detector nameplate)
A2 Meter
size
Specified meter size when ordered
(described in the Test Record)
A3 Exciting
frequency
Recommended value for the detector combined
(See 8.2.3 “Exciting Frequency.”)
A4 Flow
direction
NORMAL
A5 Password
000
B1
Indicating unit setting (main unit)
m
3
/h (If any range is specified, the same unit as that of
the range)
B2
Indicating unit setting (sub display)
% (If count rate is specified, the same unit as that of the
count rate)
B3 Custom
coefficient
0
B4
Custom unit
“ ” (7 blank spaces)
C1 Range
type
Range type specified when ordered
(If multi-range is not specified, Single range is set)
C2 to 4 Ranges 1 to 4
Values specified when ordered (described in the Test
Record. If not specified, Range 1 is the Toshiba
standard range and Ranges 2 to 4 are zeros)
C5 Hysteresis
3.0
%
Range of 1m/s or more
1.0s
Meter size
16”(400mm) or less
Range of less than 1m/s
Meter size 20” (500mm) or more
3.0s
D1 Damping
factor
For partially filled pipes detectors
10s
D2
Low cutoff value
1.0 %
D3
Output when an alarm occurs
4mA output
OFF
D4
Display low cut
For partially filled pipes detectors: ON
D5
Output low limit setting
4mA
F1
Digital Output 1
PULSE OUT (Pulse Output)
F2 to 4 Digital Outputs 2 to 4
NO USE (Not used)
F5 to 8 Digital Output 1 to 4 states
NORMAL OPEN
G1 to 2 Digital Inputs 1 to 2
NO USE (Not used)
G3 to 4 Digital inputs 1 to 2 detection levels
H LEVEL