6F8A0917

- 88 -

Changing the output low limit value setting.

The following is an example to change the output low limit setting from 4.0mA (1: 4.0mA) to
3.2mA (2: 3.2mA).

Switch operation

Display example

Description

D5: LOW LIMIT
1:4.0mA

Select D5: LOW LIMIT from the setting item selection menu.
The currently set output low limit value appears.

D5: LOW LIMIT
1:4.0mA

The cursor appears.

D5: LOW LIMIT
2:3.2mA

Press

to select 2:3.2mA.

D5: LOW LIMIT
2:3.2mA

The display blinks and the set value is displayed for

confirmation.

Press

MENU/ENT

again to write the data. Then the cursor

disappears and the display returns to the condition that the

setting value is displayed.

D5: LOW LIMIT

Returns to the menu of setting item selection.