Checking the coefficient of custom unit
Proceed as follows to check the coefficient of custom unit.
Switch operation
Display example
Description
B3: CUSTOM DATA
12340.00
Select B3: CUSTOM DATA from the setting item selection
menu.
The currently selected coefficient appears.
B3: CUSTOM DATA
Returns to the menu of setting item selection.
Changing the coefficient of custom unit
The following is an example to show how to change the coefficient from 1.00 to 1.25.
Switch operation
Display example
Description
B3: CUSTOM DATA
1.000000
Select B3: CUSTOM DATA from the setting item selection
menu.
The currently set coefficient appears.
B3: CUSTOM DATA
1.000000
The cursor appears.
Press
further to move the cursor to the desired digit to
change.
▲
B3: CUSTOM DATA
1.200000
↓
1.250000
Change the number of the digit.
To change other digit, press
to move the cursor to that digit
and change the number there. Repeat this operation until
1.250000 appears on the display.
B3: CUSTOM DATA
1.250000
The display blinks and the set value is displayed for
confirmation.
Press
MENU/ENT
again to write the data. Then the cursor
disappears and the display returns to the condition that the
setting value is displayed.
B3: CUSTOM DATA
Returns to the menu of setting item selection.