Checking the normal indicating unit
The following is an example to change the main indicating unit (UNIT1) from % to ml/s.
In the case of sub unit (UNIT2), select the function “B2: UNIT 2.”
Switch operation
Display example
Description
B1: UNIT 1
%
Select B1: UNIT 1 from the setting item selection menu.
The currently set main indicating unit (% in this example)
appears.
B1: UNIT 1
%
The cursor appears.
B1: UNIT 1
ml B
Change the volumetric unit to ml. (Note 1)
B1: UNIT 1
ml B
Move the cursor to the position of time unit.
B1: UNIT 1
ml/s
Change the time unit to s. (Note 2)
B1: UNIT 1
ml/s
The display blinks and the set value is displayed for
confirmation.
Press
MENU/ENT
again to write the data. Then the cursor
disappears and the display returns to the condition that the
setting value is displayed.
B1: UNIT 1
Returns to the menu of setting item selection.
Note 1: Volumetric unit is displayed cyclically as follows:
Note 2: Time unit is displayed cyclically loops as follows:
%
m
3
ml
RANGE
COUNT
ft/s
l
gal
bbl
m/s
pt
qt
/s
/min
/h
/d