8.5 Zero Adjustment

8.5.1 Still Water Zero Adjustment

Still water zero adjustment must be performed when the fluid in the measuring pipe of the detector
stays still.

Switch operation

Display example

Description

10.00 m/s
100.0 %

Measured value being displayed (measurement mode)

ENTER111
000

Pressing

MENU/ENT

in the measurement mode for 3 seconds

or more, the screen moves to the menu lock cancellation
screen.
* Pressing

CANCEL

in the mode selection screen, the mode

returns to the measurement mode.

ENTER111
111

Set the menu lock cancellation cord [111] by
pressing

and

.

* Pressing

CANCEL

in the mode selection screen, the mode

returns to the measurement mode.

>1:SET
2:CNT CTRL

Press

MENU/ENT

for 3 seconds or more in the measurement

mode and then the mode changes to the mode selection screen.

A1:EX CURRENT

Select 1:SET in the mode selection screen and then the menu
screen appears.

E1: ZERO ADJUST

Press

and

so that the function number becomes E1:

ZERO ADJUST.

E1: ZERO ADJUST
01.0%

The current flow rate measured value in % unit appears.
(Check that the fluid to be measured stays still.)

ADJUST READY
01.0%

ADJUST READY appears and the system is ready for
adjustment. (If “CANCEL” is pressed, Adjustment ready
condition will be cancelled.)

E1: ZERO ADJUST
* ZERO ADJUST

About 3 to 6 seconds later, the still water zero adjustment ends
and the measured flow value after adjustment appears.

E1: ZERO ADJUST
00.0%

Returns to the menu of setting item selection.

E1: ZERO ADJUST

Press

and

so that the number digit of the function

number becomes 0.

Note2

Note2