beautypg.com

Toshiba Tohsiba Electromagnetic Flowmeter Converter L5232 User Manual

Page 167

background image

6F8A0917

- 166 -

3. Other specifications

• Conductivity:

5μS/cm or more
(If detector is partially filled pipes type, conductivity is 100μS/cm or more)

• Power supply:

100 to 240VAC (allowable voltage range: 80 to 264VAC, 50/60Hz)

*1

or

24VDC (allowable voltage range: 21 to 27VDC)

*2

• Power consumption: 17W (27VA ) or less (when 7th digit of specification code is “A”)

35W (60VA ) or less (when 7th digit of specification code is “B”)
30W (50VA ) or less (when 7th digit of specification code is “F”)

*1

When the 7th digit of specification code is “B” and “F”, power supply is 100 to 120VAC

(allowable voltage range is 80 to 132VAC, 50/60Hz)

*2

When the 7th digit of specification code is “B” and “F”, 24VDC cannot be selected.

• For partially filled pipes detectors:

Fluid water level range:

Meter size 6”-12” (150-300mm) : Water level 1 1/4(30mm) to state to be filled with water

Meter size 14”-24” (350-600mm) : 10% of meter size to state to be filled with water

Note) State to be filled with water is water level 100%

Necessary straight pipe length:

Upstream side : 10D or more

Downstream side: 5D or more

Note) D is meter size.

Converter specifications

1. Input signal: Flow rate proportional signal from the detector

2. Output signal: 4 to 20mADC (current signal, load resistance 0 to 1k

Ω)

3 Digital Input/Output

• Digital Output

Digital Output 1: Output type

Transistor open collector

Number of outputs 1
Capacity 30VDC,

200mA

max.

Digital Output 2 (option: added when 9th digit of specification code is “3”) :

Output type

Solid-state contact output

Number of outputs 3
Capacity 150VDC,

150mA

max.

150VAC (peak value), 100mA max.

* For digital output function, either of the following functions can be selected.

Digital Output function
• Totalizer pulse output DO1 or DO2 can be selected

(Forward direction pulse and reverse direction pulse can be assigned independently)

In the case of DO1 Pulse rate 3.6 to 3,600,000 pulses/h
Pulse width

Settable within the range of 0.3 to 500ms

In the case of DO2 Pulse rate 3.6 to 360,000 pulses/h
Pulse width

Settable within the range of 4 to 500ms