Changing the zero offset adjustment value

Changing the zero offset adjustment value from +1.0% to

−2.5%.

Switch operation

Display example

Description

O1: MANUAL ZERO
+001.0%

Select O1: MANUAL ZERO from the setting item selection menu.
The currently set zero offset adjustment value appears.

O1: MANUAL ZERO
+001.0%

The cursor appears.
Press

further to change the cursor to the desired digit to

change.

O1: MANUAL ZERO
-001.0%

-002.5%

Change the number of the digit.
To change other digit, press

under this condition to move

the cursor to that digit and change the number there.

O1: MANUAL ZERO
-002.5%

The display blinks and the set value is displayed for

confirmation.

Press

MENU/ENT

again to write the data. Then the cursor

disappears and the display returns to the condition that the

setting value is displayed.

O1: MANUAL ZERO

Returns to the menu of setting item selection.

Note: If you try to set a value outside of the settable range, an error message of *H. OVER SPEC or *L. OVER

SPEC appears. Try to set a new value again.