Toshiba Tohsiba Electromagnetic Flowmeter Converter L5232 User Manual

Page 166

background image

6F8A0917

- 165 -

2. Measurement accuracy: (Accuracy when combined with detector)

• Combined detector: LF470 (Meter size 1/10” to 1/4” (2.5 to 6mm))

Range

Flow rate to the range (%)

1.0-3.3ft/s

(0.3-1.0m/s )

3.3-32.8ft/s

(1.0-10m/s )

0 to 50 %

± 0.8 % FS

± 0.4 % FS

50 to 100 %

± 0.8 % FS

± 0.8 % of rate

• Combined detector: LF150 (Meter size 20” to 120” (500 to 3000mm))

±0.5 % of Rate*

* This error is pulse output results calibrated under standard operating conditions at Toshiba’s

flow calibration facility.

* Individual meter’s measurement error mayvary up to ± 0.8 % of Rate at 3.28 ft/s (1.0 m/s) or

more and ± 0.4% of Rate ±0.157 inch/s (4mm/s) at 3.28 ft/s or less.

* Refer to individual calibration data for each meter’s measurement error.

• For partially filled pipes detectors (Combined detector is LF502)

± 2 % FS

Note: Accuracy is measured when detector and converter are newly combined under the basic

operating conditions in the Toshiba calibration facility.