Changing the low cutoff value

The following is an example to show how to change the low cutoff value from 1.0% to 3.0%.

Switch operation

Display example

Description

D2: LOW CUT
01.0 %

Select D2: LOW CUT from the setting item selection menu.
The currently set low cutoff value (1.0 % in this example)
appears.

D2: LOW CUT
01.0 %

The cursor appears.
Press

further to move the cursor to the desired digit to

change.

D2: LOW CUT
03.0%

Change the number of the digit.
To change another digit, press

under this condition to move

the cursor to that digit and change the number there.

D2: LOW CUT
03.0 %

The display brinks and the set value is displayed for

confirmation.

Press

MENU/ENT

again to write the data. Then the cursor

disappears and the display returns to the condition that the

setting value is displayed.

D2: LOW CUT

Returns to the menu of setting item selection.

Note: If you try to set a value exceeding 10.0%, *H. OVER SPEC.* appears and the value returns to

the previously set value. In that case, try to set a new value again.