6F8A0917

- 50 -

Example of password input (mode selection: SET and password: 123)

Switch operation

Display example

Description

INPUT PASSWORD
000

The password input screen appears from the measurement
mode. The cursor appears.
Pressing

CANCEL

, the screen returns to the mode selection

screen.

INPUT PASSWORD
100

Press

to change the value to 1.

INPUT PASSWORD
100
120
123

Change the number.
To change the number of other digit, press

to move the

cursor to that digit and press

to change the number there.

Repeat this operation until 123 appears on the screen.

INPUT PASSWORD
123

The display blinks and the set value is displayed for
conformation.
Pressing

MENU/ENT

again, the password will be written.

>1:SET
2:CNT CTRL

Regardless of whether the password input is correct or not, the
screen changes to the mode selection screen.
However, if the password input is erroneous, the setting value
cannot be changed and calibration cannot be performed.