6F8A0917

- 77 -

Changing the span value

The span value for each range is set.
The following is an example to show how to change the span value of Range 1 from 2.0m3/h to 100
l/min.

Switch operation

Display example

Description

C2: RANGE 1
2.0000 m3/h

Select C2: RANGE 1 from the setting item selection menu.
The currently set span value for Range 1 (2.0 m

3

/h in this

example) appears.

C2: RANGE 1
2.0000 m3/h

The cursor appears.
Press

further to move the cursor to the digit of volumetric

unit.

C2: RANGE 1
2000.0 l/h

33.333 l/min

Change the volumetric unit to l.

Press

to move the cursor to the digit of time unit and

likewise change the unit to min. (Note)

(The span value will be automatically changed as this unit is

changed.)

C2: RANGE 1
33.333 l/min

Move the cursor to the desired digit to change.

C2: RANGE 1
13.333 l/min

100.00 l/min

Change the number of the digit.
To change other digit, press

under this condition to move

the cursor to that digit and change the number there.

C2: RANGE 1
100.00 l/min

The display blinks and the set value is displayed for

confirmation.

Press

MENU/ENT

again to write the data. The cursor

disappears and returns to the condition that the set value is

displayed.

C2: RANGE 1

Returns to the menu of setting item selection.

Note: The items of unit are displayed cyclically as shown below:

• A combination of m/min, m/h , m/d, ft/min, ft/h and ft/d cannot be selected.

m

3

l

gal

ml

/s

/min

/h

/d

Volumetric unit

Time unit

bbl

m

ft

pt

qt