beautypg.com

Toshiba Tohsiba Electromagnetic Flowmeter Converter L5232 User Manual

Page 27

background image

6F8A0917

- 26 -

Referring to 5.2 "External Connections," connect each cable to the terminal block. Tighten the screws
of the terminal block tightly to make them securely connected. A loose connection may cause incorrect
measurement. After connecting each cable, try to pull it to check whether it has been connected
securely.

Phillips screwdriver

Terminal block

* The appropriate torque

for tightening the terminal
block screw is 1.2 N m

Figure 5.7 Connection to Terminal Block

After connecting the cables to the terminal block, take up the slack of the cables and tighten the cap
nut.

At this time, be careful that if the sheath-removed portion of the cable comes to the packing area, air
tightness may not be kept.

Cable connection

Cap nut for tightening

Cable

Packing

Correct

Because the contact
surface of the cable
and the packing is
small, air tightness
may not be kept.

Incorrect

Cable connection

Figure 5.8 Cable Tightening