Toshiba Tohsiba Electromagnetic Flowmeter Converter L5232 User Manual
Page 40

6F8A0917
- 39 -
Functions of each switch
Switch Basic
function
• Moves the mode from the measurement mode to menu display
 10.00 m/s 
 100.0 % 
 ENTER 111 
 000 
(Measurement mode)
(Menu lock cancellation screen)
• Moves the mode to the setting, calibration or measurement mode
A1:EX CURR 
 
A1:EX CURR 
 0.2100A 
(Menu display)
(Setting mode)
• Writes data in the setting mode.
B1:UNIT 1 
 % 
B1:UNIT 1 
 % 
(Data being changed)
(Data has been changed)
• Changes the numeric value or items in menu display and in the setting,
calibration or measurement mode.
C2:RANGE 1 
02.000 m/s 
C2:RANGE 1 
03.000 m/s 
(Data 0.2000m/s)
(Data 0.3000m/s)
• Starts or stops the totalizer (total value and pulse output) in the totalizer control
mode.
COUNT CTRL 
F 1000 m
3
COUNT CTRL 
F 1000 m
3
C
(Count stops)
(Count starts)
• Enables the converter to change the setting value in the setting or calibration
mode and the cursor appears.
D1: DAMPING 
 00.5 S 
D1: DAMPING 
 00.5 S 
(Setting mode)
(Enabled to change the setting value)
• Moves the cursor (digit) in menu display and in the setting mode
G1: COUNT RATE 
 1.23E-4 m
3
G1: COUNT RATE 
 1.23E-4 m
3
(Cursor at the position of “1”)
(Cursor at the position of “2”)
• Resets the totalizer (total value) in the totalizer control mode
COUNT CTRL 
F 1000 m
3
C
COUNT CTRL 
F 0 m
3
C
(Counter in operaiton)
(Resets the counter)
(*note)