Toshiba Tohsiba Electromagnetic Flowmeter Converter L5232 User Manual
Page 40
6F8A0917
- 39 -
Functions of each switch
Switch Basic
function
• Moves the mode from the measurement mode to menu display
10.00 m/s
100.0 %
ENTER 111
000
(Measurement mode)
(Menu lock cancellation screen)
• Moves the mode to the setting, calibration or measurement mode
A1:EX CURR
A1:EX CURR
0.2100A
(Menu display)
(Setting mode)
• Writes data in the setting mode.
B1:UNIT 1
%
B1:UNIT 1
%
(Data being changed)
(Data has been changed)
• Changes the numeric value or items in menu display and in the setting,
calibration or measurement mode.
C2:RANGE 1
02.000 m/s
C2:RANGE 1
03.000 m/s
(Data 0.2000m/s)
(Data 0.3000m/s)
• Starts or stops the totalizer (total value and pulse output) in the totalizer control
mode.
COUNT CTRL
F 1000 m
3
COUNT CTRL
F 1000 m
3
C
(Count stops)
(Count starts)
• Enables the converter to change the setting value in the setting or calibration
mode and the cursor appears.
D1: DAMPING
00.5 S
D1: DAMPING
00.5 S
(Setting mode)
(Enabled to change the setting value)
• Moves the cursor (digit) in menu display and in the setting mode
G1: COUNT RATE
1.23E-4 m
3
G1: COUNT RATE
1.23E-4 m
3
(Cursor at the position of “1”)
(Cursor at the position of “2”)
• Resets the totalizer (total value) in the totalizer control mode
COUNT CTRL
F 1000 m
3
C
COUNT CTRL
F 0 m
3
C
(Counter in operaiton)
(Resets the counter)
(*note)