7.3.5 Maintenance

Menu

If you enter the service code to release the protection, the maintenance specific items in the setting
mode that are not displayed normally can be accessed and the parameters that are normally prohibited
to be rewritten can be changed.

This function is mainly used by service personnel, it is not necessary for general customers to enter
this function.

If erroneously operated and the service code input screen appears, either press [CANCEL] to return to
the measurement mode or press [MENU/ENT] to go on to the setting menu. (In this case, protection
will not be released.)

Example to move to the maintenance menu

Switch operation

Display example

Description

10.00 m/s
100.0 %

Measured value being displayed (measurement mode)

ENTER111
000 V****

Pressing

MENU/ENT

in the measurement mode for 3 seconds

or more, the screen moves to the menu lock cancellation
screen.
* Pressing

CANCEL

in the mode selection screen, the mode

returns to the measurement mode.

ENTER111
111 V****

Set the menu lock cancellation cord [111] by
pressing

and

.

* Pressing

CANCEL

in the mode selection screen, the mode

returns to the measurement mode.

>1:SET
2:CNT CTRL

Pressing

MENU/ENT

for 3 seconds or more in the

measurement mode, the screen moves to the mode selection
screen.

INPUT SRV_CODE
000

Select 1:SET in the mode selection screen and press

MENU/ENT

for 5 seconds or more, then the screen appears

that is used to enter the service code to release the protection.

Pressing

CANCEL

in the mode selection screen brings you

back to the measurement mode.

A1:EX CURR

Enter the service code and if it is correct, the maintenance
specific items in the setting menu can be accessed.
Pressing

CANCEL

in the service code input screen brings you

back to the mode selection screen.

Note1: There is not this screen to the converter before serial No.072320999.

There is not software version (V****) to the converter before version V0109.

Note1

Note1