Self-diagnosis error

The converter performs an operation check immediately after power is turned on and if an error is
found, a self-diagnosis error is generated.

If more than one error occurs at the same time, the contents of those errors will be displayed
cyclically. The diagnosis about excitation is carried out using the internal ADC function. Therefore,
if an error of No.6 internal ADC occurs, Excitation wire of No.4 and Excitation circuit of No.5
cannot be diagnosed correctly.

In addition, since the entire checking system is based on the CPU, if the CPU fails, accurate
diagnosis and indication cannot be obtained.

No. LCD

display

Description

Corrective

action

1

* ROM ERROR *

ROM error occurred

2

* RAM ERROR *

RAM error occurred

3

PARAMETER
FAIL

An error occurred in the internal
parameters in the memory

Internal parts or printed-circuit board must be
repaired or replaced.
Contact your nearest Toshiba representative.

4

EX. CURR. OPEN

Excitation wires are not
connected.

Connect the excitation wires correctly.

5

EX. CURR. ERROR

An error occurred in the
excitation circuit.

6

ADC. ERROR

An error occurred in the ADC
circuit

Internal parts or printed-circuit board must be
repaired or replaced.
Contact your nearest Toshiba representative.

7

INVALID TOTAL

Totalized data in the memory
was destroyed caused by the
entry of noise.
(No message appears if the data
is not displayed in an indicating
unit of totalization.)

The error message disappears if you press the
reset key to clear the totalized data.

Note 1: Diagnosis for No. 1, No.2 and No. 3 is performed only immediately after power is turned on. If any of

these errors is indicated, the converter does not start measurement operation. If these errors occur after
power is turned on, the converter cannot detect these errors. In addition, if this happen, the indication
and output of the converter may become uncertain.

Note 2: Errors No. 4 to No. 6 may not be detected even if those errors cause accuracy problems because of the

characteristic variations in components constituting the circuits used for judgment standard to
determine these errors.

Note 3: CPU error cannot be detected. If the CPU stops, the watchdog timer resets the hardware and the

flowmeter starts again from the initial power-on condition. Depending on the condition of CPU error,
the indication and output of the converter may become uncertain.