9.2 Converter Check / Calibration

9.2.1 0% Flow Rate Calibration

(Zero Calibration)

Using Mag-Prover internal calibration circuit, 0% flow rate (hereafter called zero point) calibration
can be performed.

Checking the zero point

Switch operation

Display example

Description

P1:FLOW CAL 0
0.0 %

Select P1:FLOW CAL 0 from the calibration item selection
menu.
Zero point using the simulation input appears.

P1:FLOW CAL 0

Returns to the menu of calibration item selection.

Calibrating the zero point

Switch operation

Display example

Description

P1:FLOW CAL 0
0.1 %

Select P1:FLOW CAL 0 from the calibration item selection
menu.
Zero point using the simulation input appears.

ADJUST READY
0.1 %

Press

. Then the display shown left appears and the system

is ready for zero calibration. (Note 1)

P1:FLOW CAL 0%
* CAL. 0% ADJ.

The display shown left appears and zero calibration starts.

P1:FLOW CAL 0
0.0 %

About 3 to 6 seconds later (Note 2), the zero calibration ends
and a new zero point appears.

P1:FLOW CAL 0%

Returns to the menu of calibration item selection.

Note 1: To cancel the adjustment from the condition of ADJUST READY, press

CANCEL

. The display returns

to the condition where zero point using the simulation input is displayed.

Note 2: Calibration time differs depending on the exciting frequency.