Source]:lfoutput:sweptsine:frequency:stop:step, Source]:lfoutput:sweptsine:level – RIGOL DSG3000 Series User Manual
Page 74

RIGOL
2-54
DSG3000 Programming Guide
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STOP:STEP
Syntax
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STOP:STEP
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STOP:STEP?
Description
Set the stop frequency step of sweep-sine.
Query the stop frequency step of sweep-sine.
Parameter
Name
Type
Range
Default
Real
100mHz to 500kHz
10kHz
Explanation
When
3000. Besides,
3kHz.
The default unit of the return value is Hz.
After setting the stop frequency step, you can use the up/down direction keys
or knob to modify the stop frequency at the current step. At this point, you can
query or set the stop frequency using the
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STOP
Return Format
The query returns the stop frequency step of sweep-sine, for example, 3000.
Example
:LFO:SWEP:FREQ:STOP:STEP 3kHz
:LFO:SWEP:FREQ:STOP:STEP?
Related
Command
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STOP
[:SOURce]:LFOutput:SWEPtsine:LEVel
Syntax
[:SOURce]:LFOutput:SWEPtsine:LEVel
[:SOURce]:LFOutput:SWEPtsine:LEVel?
Description
Set the amplitude of sweep-sine.
Query the amplitude of sweep-sine.
Parameter
Name
Type
Range
Default
Real
1mV to 3V
1V
Explanation
When
can also be set in "Number + Unit" form, for example, 2V.
The default unit of the return value is V.
After setting the amplitude of sweep-sine, you can use the up/down direction
keys or knob to modify the amplitude at the current step. At this point, you
can query or set the current step using the
[:SOURce]:LFOutput:SWEPtsine:LEVel:STEP
Return Format
The query returns the amplitude of sweep-sine, for example, 2.00.
Example
:LFO:SWEP:LEV 2
:LFO:SWEP:LEV?
Related
Command
[:SOURce]:LFOutput:SWEPtsine:LEVel:STEP