Source]:sweep:reset[:all, Source]:sweep:state – RIGOL DSG3000 Series User Manual
Page 108

RIGOL
2-88
DSG3000 Programming Guide
[:SOURce]:SWEep:RESet[:ALL]
Syntax
[:SOURce]:SWEep:RESet[:ALL]
Description
Reset all the sweeps to the start point.
Explanation
If the current sweep direction is "Fwd", the instrument will stop the current sweep
and sweep from the start frequency or start level after sending this command.
If the current sweep direction is "Down", the instrument will stop the current sweep
and sweep from the stop frequency or stop level after sending this commsnd.
Related
Command
[:SOURce]:SWEep:STATe
Syntax
[:SOURce]:SWEep:STATe OFF|FREQuency|LEVel[,FREQuency]
[:SOURce]:SWEep:STATe?
Description
Set the sweep state.
Query the sweep state.
Parameter
Name
Type
Range
Default
OFF|FREQuency|
LEVel[,FREQuency]
Discrete
OFF|FREQuency|LEVel|
LEVel,FREQuency
OFF
Explanation
OFF: turn off the sweep function.
FREQuency: enable the frequency sweep function.
LEVel: enable the amplitude sweep function.
LEVel,FREQuency: enable the frequency and amplitude sweep functions at the
same time.
Return Format
The query returns the sweep state, for example, FREQ.
Example
:SWE:STAT FREQ
:SWE:STAT?