Source]:frequency:offset – RIGOL DSG3000 Series User Manual
Page 47

RIGOL
DSG3000 Programming Guide
2-27
[:SOURce]:FREQuency:OFFSet
Syntax
[:SOURce]:FREQuency:OFFSet
[:SOURce]:FREQuency:OFFSet?
Description
Set the RF frequency offset.
Query the RF frequency offset.
Parameter
Name
Type
Range
Default
Real
-20GHz to 20GHz
0Hz
Explanation
When
20000. In addition,
example, 20kHz.
The default unit of the return value is Hz.
After the RF frequency offset is set, you can use the up/down direction keys or
knob to modify the offset at the current step. You can query or set the current
[:SOURce]:FREQuency:OFFSet:STEP
When the frequency offset is 0 Hz, the frequency displayed in the interface
(namely the setting frequency) is equal to the actual output frequency.
When the frequency offset is not 0 Hz, the output frequency is determined by
the setting frequency and frequency offset. The three parameters satisfy the
equation: setting frequency (display frequency) = output frequency +
frequency offset. You can use the
and set the RF frequency.
Return Format
The query returns the RF frequency offset, for example, 20000.
Example
:FREQ:OFFS 20kHz
:FREQ:OFFS?
Related
Commands
[:SOURce]:FREQuency:OFFSet:STEP
[:SOURce]:FREQuency