Source]:level – RIGOL DSG3000 Series User Manual
Page 60

RIGOL
2-40
DSG3000 Programming Guide
[:SOURce]:LEVel
Syntax
[:SOURce]:LEVel
[:SOURce]:LEVel?
Description
Set the RF output amplitude.
Query the RF output amplitude.
Parameter
Name
Type
Range
Default
Real
-140dBm to 20dBm
-140dBm
Explanation
When
2. In addition,
2dBm. At this point, the amplitude displayed in the interface is related to the
setting of Level Unit:
—
When the level unit is "dBm", 2.00dBm is displayed;
—
When the level unit is "dBmV", 48.99dBmV is displayed;
—
When the level unit is "dBuV", 108.99dBuV is displayed;
—
When the level unit is "Volts", 281.50mV is displayed;
—
When the level unit is "Watts", 1.58mW is displayed.
The default unit of the return value is dBm.
After the RF amplitude is set, you can use the up/down direction keys or knob
to modify the amplitude at the current step. You can query or set the current
When the amplitude offset is 0 dB, the amplitude displayed in the interface
(namely the setting amplitude) is equal to the actual output amplitude.
When the amplitude offset is not 0 dB, the output amplitude is determined by
the setting amplitude and amplitude offset. The three parameters satisfy the
equation: setting amplitude (display amplitude) = output amplitude +
amplitude offset. You can use the
and set the RF amplitude offset.
Return Format
The query returns the RF amplitude, for example, 2.00.
Example
:LEV 2dBm /*Set the RF amplitude to 2dBm*/
:LEV? /*Query the RF amplitude and the query returns 2.00*/
Related
Commands
[:SOURce]:LEVel:STEP
[:SOURce]:LEVel:OFFSet