[:SOURce]:SWEep:SWEep:TRIGger:TYPE
Syntax
[:SOURce]:SWEep:SWEep:TRIGger:TYPE AUTO|KEY|BUS|EXT
[:SOURce]:SWEep:SWEep:TRIGger:TYPE?
Description
Set the trigger mode of the sweep period.
Query the trigger mode of the sweep period.
Parameter
Name
Type
Range
Default
AUTO|KEY|BUS|EXT
Discrete
AUTO|KEY|BUS|EXT
AUTO
Explanation
AUTO: select "Auto" trigger mode. If the sweep mode is set to "Cont", the
instrument will start sweeping when a sweep type is selected. If the sweep
mode is set to "Single", the instrument will start a sweep and then stop after
command is sent and when the single sweep
condition is met.
KEY: select "Key" trigger mode. If the sweep mode is set to "Cont", the
instrument starts a sweep each time Trigger at the front panel is pressed. If
the sweep mode is set to "Single", the instrument starts a sweep and then stops
each time Trigger at the front panel is pressed after the
command is sent and when the single sweep
condition is met.
BUS: select "Bus" trigger mode. If the sweep mode is set to "Cont", the
instrument starts a sweep each time the
command is sent. If the sweep mode is set
to "Single", the instrument starts a sweep and then stops each time the
command is sent and when the single sweep
condition is met.
EXT: select "Ext" trigger mode. The RF signal generator receives the trigger
signal input from the [TRIGGER IN] connector at the rear panel. If the sweep
mode is "Cont", the instrument starts a sweep each time a TTL pulse signal with
the specified polarity is received. If the sweep mode is set to "Single", the
instrument starts a sweep and then stops each time a TTL pulse signal with the
specified polarity is received after the
sent and when the single sweep condition is met.
Note: The above explanations are only valid when the trigger mode of each
sweep point within the sweep period is met.
When executing the sweep operation, the priority of the required conditions is:
single sweep trigger mode point trigger mode.
Return Format
The query returns the trigger mode of the sweep, for example, AUTO.
Example
:SWE:SWE:TRIG:TYPE AUTO
:SWE:SWE:TRIG:TYPE?
Related
Commands
[:SOURce]:SWEep:EXECute
[:SOURce]:SWEep:MODE
[:SOURce]:SWEep:POINt:TRIGger:TYPE
*TRG
:TRIGger[:SWEep][:IMMediate]