Source]:sweep:step:dwell, Source]:sweep:step:dwell:step – RIGOL DSG3000 Series User Manual
Page 109

RIGOL
DSG3000 Programming Guide
2-89
[:SOURce]:SWEep:STEP:DWELl
Syntax
[:SOURce]:SWEep:STEP:DWELl
[:SOURce]:SWEep:STEP:DWELl?
Description
Set the dwell time of step sweep.
Query the dwell time of step sweep.
Parameter
Name
Type
Range
Default
Real
20ms to 100s
500ms
Explanation
When
can also be set in "Number + Unit" form, for example, 3000ms.
The default unit of the return value is s.
After setting the dwell time, you can use the up/down direction keys or knob to
modify the dwell time at the current step. At this point, you can query or set the
current step using the
[:SOURce]:SWEep:STEP:DWELl:STEP
Return Format
The query returns the dwell time of step sweep, for example, 3.
Example
:SWE:STEP:DWEL 3 /*Set the dwell time of step sweep to 3s*/
:SWE:STEP:DWEL? /*The query returns 3*/
Related
Command
[:SOURce]:SWEep:STEP:DWELl:STEP
[:SOURce]:SWEep:STEP:DWELl:STEP
Syntax
[:SOURce]:SWEep:STEP:DWELl:STEP
[:SOURce]:SWEep:STEP:DWELl:STEP?
Description
Set the dwell time step.
Query the dwell time step.
Parameter
Name
Type
Range
Default
Real
10ms to 10s
10ms
Explanation
When
can also be set in "Number + Unit" form, for example, 3000ms.
The default unit of the return value is s.
After setting the dwell time step, you can use the up/down direction keys or
knob to modify the dwell time at the current step. At this point, you can query
or set the dwell time using the
Return Format
The query returns the dweel time step, for example, 3.
Example
:SWE:STEP:DWEL:STEP 3 /*Set the dwell time step to 3s*/
:SWE:STEP:DWEL:STEP? /*The query returns 3*/
Related
Command