Source]:level:offset:step, Source]:level:sh:active – RIGOL DSG3000 Series User Manual
Page 66

RIGOL
2-46
DSG3000 Programming Guide
[:SOURce]:LEVel:OFFSet:STEP
Syntax
[:SOURce]:LEVel:OFFSet:STEP
[:SOURce]:LEVel:OFFSet:STEP?
Description
Set the step of the RF amplitude offset.
Query the step of the RF amplitude offset.
Parameter
Name
Type
Range
Default
Real
0.01dB to 200dB
10dB
Explanation
When
The default unit of the return value is dB.
After setting the step of the amplitude offset, you can use the up/down
direction keys or knob to modify the amplitude offset at the current step. At
this point, you can query or set the amplitude offset using the
Return Format
The query returns the step of the RF amplitude offset, for example, 20.00.
Example
:LEV:OFFS:STEP 20
:LEV:OFFS:STEP?
Related
Command
[:SOURce]:LEVel:SH:ACTIve
Syntax
[:SOURce]:LEVel:SH:ACTIve
Description
Execute a sample hold operation immediately.
Explanation
This command is valid when the ALC function is disabled.
Related
Command