Source]:level:alc:mode, Source]:level:attenuation – RIGOL DSG3000 Series User Manual
Page 61

RIGOL
DSG3000 Programming Guide
2-41
[:SOURce]:LEVel:ALC:MODE
Syntax
[:SOURce]:LEVel:ALC:MODE OFF|ON|AUTO
[:SOURce]:LEVel:ALC:MODE?
Description
Set the working state of the ALC function.
Query the working state of the ALC function.
Parameter
Name
Type
Range
Default
OFF|ON|AUTO Discrete
OFF|ON|AUTO
AUTO
Explanation
OFF: turn off the ALC function. At this point, the instrument switches to
"Sample & Hold" state.
ON: turn on the ALC function.
AUTO: the current working state of the ALC function is automatic.
Return Format
The query returns OFF, ON or AUTO.
Example
:LEV:ALC:MODE OFF /*Turn off the ALC function*/
:LEV:ALC:MODE? /*The query returns OFF*/
Related
Commands
[:SOURce]:LEVel:SH:ACTIve
[:SOURce]:LEVel:SH:MODE
[:SOURce]:LEVel:ATTenuation
Syntax
[:SOURce]:LEVel:ATTenuation
[:SOURce]:LEVel:ATTenuation?
Description
Set the attenuation of the RF output amplitude.
Query the attenuation of the RF output amplitude.
Parameter
Name
Type
Range
Default
Integer
0dB to 125dB
125dB
Explanation
When
The attenuation can only be set to multiples of 5.
The default unit of the return value is dB.
After the amplitude attenuation is set, you can use the up/down direction keys
or knob to modify the attenuation at the current step. You can query or set the
[:SOURce]:LEVel:ATTenuation:STEP
You can set the attenuation only when the "FIXed" attenuation mode is
[:SOURce]:LEVel:ATTenuation:MODE
mode, this command is invalid.
Return Format
The query returns the RF amplitude attenuation, for example, 20.00.
Example
:LEV:ATT 20dB
:LEV:ATT?
Related
Commands
[:SOURce]:LEVel:ATTenuation:STEP
[:SOURce]:LEVel:ATTenuation:MODE