Source]:pulm:width – RIGOL DSG3000 Series User Manual
Page 100

RIGOL
2-80
DSG3000 Programming Guide
[:SOURce]:PULM:WIDTh
Syntax
[:SOURce]:PULM:WIDTh
[:SOURce]:PULM:WIDTh?
Description
Set the width of the pulse modulating signal.
Query the width of the pulse modulating signal.
Parameter
Name
Type
Range
Default
Real
10ns to 170s-10ns
500us
Explanation
When
can also be set in "Number + Unit" form, for example, 2000ms.
The default unit of the return value is s.
When the modulation source is set to "Int", you can use this command to set
the width of single pulse if the pulse type is set to "Single" and you can use this
command to set the width of #1 pulse if the pulse type is set to "Double".
Otherwise, this command is invalid.
After setting the pulse width, you can use the up/down direction keys or knob
to modify the pulse width at the current step. At this point, you can query or set
the current step using the
The single pulse width is limited by the minimum pulse width and pulse period
and the three fulfill the following relations.
Pulse Width ≥ Minimum Pulse Width
Pulse Width ≤ Pulse Period – 10ns
The width of #1 pulse of double pulse is limited by the minimum pulse width
and delay.
Pulse Width ≥ Minimum Pulse Width
Pulse Width ≤ Delay – 10 ns
Return Format
The query returns the width of the pulse modulating signal, for example, 2.
Example
:PULM:WIDT 2
:PULM:WIDT?
Related
Commands
[:SOURce]:PULM
[:SOURce]:PULM:MODE
[:SOURce]:PULM:PERiod
[:SOURce]:PULM:SOURce
[:SOURce]:PULM:WIDTh:STEP