Source]:level:offset – RIGOL DSG3000 Series User Manual
Page 65

RIGOL
DSG3000 Programming Guide
2-45
[:SOURce]:LEVel:OFFSet
Syntax
[:SOURce]:LEVel:OFFSet
[:SOURce]:LEVel:OFFSet?
Description
Set the amplitude offset of the RF output amplitude.
Query the amplitude offset of the RF output amplitude.
Parameter
Name
Type
Range
Default
Real
-200dB to 200dB
0dB
Explanation
When
The default unit of the return value is dB.
After setting the amplitude offset, you can use the up/down direction keys or
knob to modify the amplitude offset at the current step. At this point, you can
query or set the current step using the
command.
When the amplitude offset is 0 dB, the display amplitude (namely the setting
amplitude) is equal to the actual output amplitude.
When the amplitude offset is not 0 dB, the output amplitude is determined by
the setting amplitude and amplitude offset. The three parameters satisfy the
equation: setting amplitude (display amplitude) = output amplitude +
amplitude offset. You can query or set the RF amplitude using the
Return Format
The query returns the offset of the RF output amplitude, for example, 10.00.
Example
:LEV:OFFS 10
:LEV:OFFS?
Related
Commands
[:SOURce]:LEVel:OFFSet:STEP
[:SOURce]:LEVel