Rst -4, Sre -4, Stb? -4 – RIGOL DSG3000 Series User Manual
Page 24: Trg -4

RIGOL
2-4
DSG3000 Programming Guide
*RST
Syntax
*RST
Description
Restore the RF signal generator to the preset state (factory or user).
Related
Command
*SRE
Syntax
*SRE
*SRE?
Description
Set the enable register for the status byte register.
Query the enable register for the status byte register.
Parameter
Name
Type
Range
Default
Integer
0 to 255
0
Explanation
The bit 0 and bit 1 of the status byte register are not used and are always treated as
0, therefore, the range of
binary numbers ranging from 00000000 (0 in decimal) to 11111111 (255 in decimal)
and of which the bit 0 and bit 1 are 0.
Return Format
The query returns an integer which equals the sum of the weights of all the bits that
have already been set in the register.
Example
*SRE 16 /*Enable the bit 4 (16 in decimal) of the status byte register */
*SRE? /*The query returns 16*/
*STB?
Syntax
*STB?
Description
Query the event register for the status byte register. The value of the status byte
register is set to 0 after this command is executed.
Explanation
The bit 0 and bit 1 of the status byte register are not used and are always treated as
0. The query returns the decimal numbers corresponding to the binary numbers
ranging from 00000000 (0 in decimal) to 11111111 (255 in decimal) and of which the
bit 0 and bit 1 are 0.
Return Format
The query returns an integer which equals the sum of the weights of all the bits that
have already been set in the register.
*TRG
Syntax
*TRG
Description
Trigger a pulse modulation, RF sweep or LF sweep-sine immediately.
Related
Commands
:TRIGger:LFOutput[:IMMediate]
:TRIGger:PULSe[:IMMediate]
:TRIGger[:SWEep][:IMMediate]