Source]:correction:flatness:pair, Source]:correction:flatness[:state, Source]:fm command subsystem – RIGOL DSG3000 Series User Manual
Page 41: Source]:fm command subsystem -21

RIGOL
DSG3000 Programming Guide
2-21
[:SOURce]:CORRection:FLATness:PAIR
Syntax
[:SOURce]:CORRection:FLATness:PAIR
Description
Set the correction point pair in the flatness table.
Parameter
Name
Type
Range
Default
Real
9kHz to 6GHz
6GHz
Real
-100dB to 100dB
NULL
Explanation
You can use this command to edit the frequency and amplitude of the row
currently selected in the flatness table.
Example
:CORR:FLAT:PAIR 2.5M,10 /*Set the frequency of the row currently selected to
2.5MHz and the amplitude to 10dB*/
[:SOURce]:CORRection:FLATness[:STATe]
Syntax
[:SOURce]:CORRection:FLATness[:STATe] ON|OFF|1|0
[:SOURce]:CORRection:FLATness[:STATe]?
Description
Turn on or off the flatness correction switch.
Query the state of the flatness correction switch.
Parameter
Name
Type
Range
Default
ON|OFF|1|0
Bool
ON|OFF|1|0
OFF
Return Format
The query returns 1 or 0.
Example
:CORR:FLAT ON /*Turn on the flatness correction function*/
:CORR:FLAT? /*The query returns 1*/
[:SOURce]:FM Command Subsystem
Command List:
[:SOURce]:FM:DEViation:STEP[:INCRement]
[:SOURce]:FM:FREQuency:STEP[:INCRement]