Source]:level:sh:mode, Source]:level:step – RIGOL DSG3000 Series User Manual
Page 67

RIGOL
DSG3000 Programming Guide
2-47
[:SOURce]:LEVel:SH:MODE
Syntax
[:SOURce]:LEVel:SH:MODE AUTO|MANUal
[:SOURce]:LEVel:SH:MODE?
Description
Set the current sample hold mode.
Query the current sample hold mode.
Parameter
Name
Type
Range
Default
AUTO|MANUal
Discrete
AUTO|MANUal
AUTO
Explanation
A sample hold operation is performed automatically each time the RF
amplitude is set.
This command is valid when the ALC function is disabled.
Return Format
The query returns AUTO or MANUAL.
Example
:LEV:SH:MODE MANU
:LEV:SH:MODE?
Related
Commands
[:SOURce]:LEVel:ALC:MODE
[:SOURce]:LEVel:SH:ACTIve
[:SOURce]:LEVel:STEP
Syntax
[:SOURce]:LEVel:STEP
[:SOURce]:LEVel:STEP?
Description
Set the RF output amplitude step.
Query the RF output amplitude step.
Parameter
Name
Type
Range
Default
Real
0.01dB to 100dB
10dB
Explanation
When
The default unit of the return value is dB.
After setting the output amplitude step, you can use the up/down direction
keys or knob to modify the output amplitude at the current step. At this point,
you can query or set the output amplitude using the
command.
Return Format
The query returns the RF output amplitude step, for example, 20.00.
Example
:LEV:STEP 20
:LEV:STEP?
Related
Command