AMETEK BPS Series Programming Manual User Manual
Page 96

AMETEK Programmable Power
BPS / MX / RS Series SCPI Programming Manual
96
Manual P/N 7003-961 Rev AA
4.17.8
List:Voltage
LIST:VOLTage
Phase Selectable
This command specifies the output voltage points in a list. The voltage points are given in
the command parameters, which are separated by commas. The order in which the points
are entered determines the sequence in which the list will be output when a list transient is
triggered.
Command Syntax
[SOURce:]LIST:VOLTage[:LEVel]
Parameters
Value depends on the voltage range and the voltage mode
Unit
V (rms voltage)
Examples LIST:VOLT
2.0,2.5,3.0
LIST:VOLT
MAX,2.5,MIN
Query Syntax
[SOURce:]LIST:VOLTage[:LEVel]?
Returned Parameters
Related Commands
LIST:VOLT:POIN?
LIST:COUN
LIST:DWELLIST:STEP
LIST:SHAP
LIST:VOLT:OFFS
LIST:VOLTage:POINts?
Phase Selectable
This query returns the number of points specified in LIST:VOLT. Note that it returns only the
total number of points, not the point values.
Query Syntax
[SOURce:]LIST:VOLTage:POINts?
Returned Parameters
Example LIST:VOLT:POIN?
Related Commands
LIST:VOLT
LIST:VOLTage:SLEW
Phase Selectable
This command specifies the output offset slew points in a list. The slew points are given in
the command parameters, which are separated by commas. The order in which the points
are entered determines the sequence in which the list will be output when a list transient is
triggered. Changing list data while a subsystem is in list mode generates an implied ABORt.
Command Syntax
[SOURce:]LIST:VOLTage:SLEW
Parameters
0.1 to 1E9 | MAX
Unit
V/S (volts per second)
Example
LIST:VOLT:SLEW 10, 1E2, MAX
Query Syntax
[SOURce:]LIST:VOLTage:SLEW?
Returned Parameters
Related Commands
LIST:VOLT:SLEW:POIN?
LIST:COUN
LIST:DWEL LIST:STEP