AMETEK BPS Series Programming Manual User Manual
Page 45

BPS / MX / RS Series SCPI Programming Manual
AMETEK Programmable Power
Manual P/N 7003-961 Rev. AA
45
CALibrate:MEASure:VOLTage:TEMPerature:DC[:FSCale]
This command initiates the calibration of the DC voltage measurement at full scale and at
elevated temperature.
Command Syntax
CALibrate:MEASure:VOLTage:TEMPerature:DC[:FSCale]
Parameters
Examples CAL:MEAS:VOLT:TEMP
120
Query Syntax
CALibrate:MEASure:VOLTage:TEMPerature:DC?
Returned Parameters
Related Commands
CALibrate:MEASure:VOLTage[:AMBient]:DC[:FSCale]
CALibrate:MEASure:VOLTage:TEMPerature:DC:ZERO
This command initiates the offset adjustment of the rms voltage measurement at elevated
temperature.
Command Syntax
CALibrate:MEASure:VOLTage:TEMPerature:DC:ZERO
Parameters
Examples CAL:MEAS:VOLT:TEMP:DC:ZERO
0
Query Syntax
CALibrate:MEASure:VOLT:DC:ZERO?
Returned Parameters
Related Commands
CALibrate:MEASure:VOLTage[:AMBient]:DC:ZERO