BPS / MX / RS Series SCPI Programming Manual

AMETEK Programmable Power

Manual P/N 7003-961 Rev. AA

45

CALibrate:MEASure:VOLTage:TEMPerature:DC[:FSCale]

This command initiates the calibration of the DC voltage measurement at full scale and at
elevated temperature.

Command Syntax

CALibrate:MEASure:VOLTage:TEMPerature:DC[:FSCale]

Parameters

(actual DC output voltage measured with external device)

Examples CAL:MEAS:VOLT:TEMP

120

Query Syntax

CALibrate:MEASure:VOLTage:TEMPerature:DC?

Returned Parameters

(value range -1000 to +1000)

Related Commands

CALibrate:MEASure:VOLTage[:AMBient]:DC[:FSCale]

CALibrate:MEASure:VOLTage:TEMPerature:DC:ZERO

This command initiates the offset adjustment of the rms voltage measurement at elevated
temperature.

Command Syntax

CALibrate:MEASure:VOLTage:TEMPerature:DC:ZERO

Parameters

(0 or desired offset value)

Examples CAL:MEAS:VOLT:TEMP:DC:ZERO

0

Query Syntax

CALibrate:MEASure:VOLT:DC:ZERO?

Returned Parameters

(value range -20 to +20)

Related Commands

CALibrate:MEASure:VOLTage[:AMBient]:DC:ZERO