4.17.3

List - Frequency

LIST:FREQuency

This command sets the sequence of frequency list points. The frequency points are given in
the command parameters, which are separated by commas. The order in which the points
are entered determines the sequence in which they are output when a list is triggered.

Command Syntax

[SOURce:]LIST:FREQuency[:LEVel],

Parameters

Values limited to the frequency range

Unit HZ

(Hertz)

Examples LIST:FREQ

60,65,70

Query Syntax

[SOURce:]LIST:FREQ?

Returned Parameters

Related Commands

LIST:FREQ:POIN?

LIST:COUN

LIST:DWEL

LIST:STEP LIST:FREQ:SLEW

LIST:FREQuency:POINts?

This query returns the number of points specified in LIST:FREQuency. Note that it returns
only the total number of points, not the point values.

Query Syntax

[SOURce:]LIST:FREQ[:LEVel]:POINts?

Returned Parameters

Example LIST:FREQ:POIN?
Related Commands

LIST:FREQ

LIST:FREQuency:SLEW

This command sets the sequence of frequency slew list points. The frequency points are
given in the command parameters, which are separated by commas.The order in which the
points are entered determines the sequence in which they are output when a list is triggered.

Command Syntax

[SOURce:]LIST:FREQuency:SLEW,

Parameters

0.01 to 1E9

⏐MAXimum

Unit

HZ (Hertz) per second

Examples

LIST:FREQ:SLEW 10, 1E2, MAX

Query Syntax

[SOURce:]LIST:FREQ:SLEW?

Returned Parameters

Related Commands

LIST:FREQ:SLEW:POIN?

LIST:COUN

LIST:DWEL LIST:STEP LIST:FREQ

LIST:FREQuency:SLEW:POINts?

This query returns the number of points specified in LIST:FREQuency:SLEW. Note that it
returns only the total number of points, not the point values.

Query Syntax

[SOURce:]LIST:FREQ:SLEW:POINts?

Returned Parameters

Example LIST:FREQ:SLEW:POIN?
Related Commands

LIST:FREQ:SLEW