5.10

*SAV

This command stores the present state of the AC/DC power source to a specified location in
memory. Up to 8 states (including the transient list) can be stored in nonvolatile memory.

Command Syntax

*SAV

Parameters

0 through 7

Related Commands

PSC

*RCL *RST

5.11

*SRE

This command sets the condition of the Service Request Enable Register. This register
determines which bits from the Status Byte Register (see *STB for its bit configuration) are
allowed to set the Master Status Summary (MSS) bit and the Request for Service (RQS)
summary bit. A 1 in any Service Request Enable Register bit position enables the
corresponding Status Byte Register bit and all such enabled bits then are logically ORed to
cause Bit 6 of the Status Byte Register to be set. See paragraph 7.5 for more details
concerning this process.

When the IEEE-488 BUS controller conducts a serial poll in response to SRQ, the RQS bit is
cleared, but the MSS bit is not. When *SRE is cleared (by programming it with 0), the source
cannot generate an SRQ to the controller.

Command Syntax

*SRE

Parameters

0 to 255

Default Value

0 (see *PSC command)

Example *SRE

255

Query Syntax

*SRE?

Returned Paramters

(Register binary value)

Related Commands

*ESE *ESR