BPS / MX / RS Series SCPI Programming Manual
176
Manual P/N 7003-961 Rev AA
IEC411:VARiants:HOLD:TIME
This command will set the variant hold time in seconds. Refer to Figure 9-13 in the User
Manual.
Command Syntax
IEC411:VARiants: HOLD:TIME
Parameters
Examples
IEC411:VAR: HOLD:TIME 0.5
Query Syntax
IEC411:VAR: HOLD:TIME?
Returned Parameters
IEC411:VARiants:HOLD:VOLTage[:PERCent]
This command will set the variant level as a percentage of the nominal volts for the duration
of the hold time.
Command Syntax
IEC411:VARiants:HOLD:VOLTage[:PERCent]
Parameters
Examples
IEC411:VAR: HOLD:VOLT 70
Query Syntax
IEC411:VAR: HOLD:VOLT?
Returned Parameters
IEC411:VARiants:FALL[:TIME]
This command will set the variant fall time in seconds. Refer to Figure 9-13 in the User
Manual.
Command Syntax
IEC411:VARiants:FALL[:TIME]
Parameters
Examples
IEC411:VAR:FALL 0.5
Query Syntax
IEC411:VAR:FALL?
Returned Parameters
IEC411:VARiants:RISE[:TIME}
This command will set the variant rise time in seconds. Refer to Figure 9-13 in the User
Manual.
Command Syntax
IEC411:VARiants:RISE:TIME
Parameters
Examples IEC411:VAR:RISE:TIME
0.25
Query Syntax
IEC411:VAR:RISE:TIME?
Returned Parameters