AMETEK BPS Series Programming Manual User Manual
Page 61

BPS / MX / RS Series SCPI Programming Manual
AMETEK Programmable Power
Manual P/N 7003-961 Rev. AA
61
Query Syntax
MEASure:ARRay:VOLTage:HARMonic[:AMPLitude]? [
FETCh:ARRay:VOLTage:HARMonic[:AMPLitude]? [
Parameters None
Examples MEAS:ARR:VOLT:HARM?
FETC:ARR:VOLT:HARM?
Returned Parameters
51 NR2 values
Related Commands
INST:NSEL
MEASure:ARRay:VOLTage:HARMonic:PHASe? [
FETCh:ARRay:VOLTage:HARMonic:PHASe? [
Phase Selectable
These queries return an array of harmonic phases of output voltage in degrees, referenced
to the positive zero crossing of the fundamental component. The fundamental component
will return a value relative to the fundamental voltage for phase A. Phase A will return a zero
value.
The first value returned is the dc component (always returned as 0 degrees phase); the
second value is the fundamental frequency, and so on up to the 50th harmonic. Harmonic
orders can be measured up to the fundamental measurement bandwidth of the
measurement system:
MX Series I:
25.6 microseconds for single-phase mode and 76.8 microseconds for three-
phase mode.
MX/RS Series II: 10.4 microseconds for single-phase mode and 31.2 microseconds for
three-phase mode.
Thus the maximum harmonic that can be measured is dependent on the output frequency.
Any harmonics that represent frequencies greater than the above frequencies are returned
as 0.
Query Syntax
MEASure:ARRay:VOLTage:HARMonic:PHASe?
FETCh:ARRay:VOLTage:HARMonic:PHASe?
Parameters None
Examples MEAS:ARR:VOLTage:HARM:PHAS?
30
FETC:ARR:VOLTage:HARM:PHAS?
Returned Parameters
31 NR2 values
Related Commands
INST:NSEL