AMETEK BPS Series Programming Manual User Manual
Page 74

AMETEK Programmable Power
BPS / MX / RS Series SCPI Programming Manual
74
Manual P/N 7003-961 Rev AA
OUTPut:IMMediate
This command opens the output relay without first programming the voltage down if the –
SNK option is installed and the MX/RS/BPS is in REGENERATIVE mode (REG:STAT ON).
If not, this commands works exactly the same as the OUTP[:STAT] command. This mode
may be used to simulate line loss in inverter applications. The query form returns the output
state.
Command Syntax
OUTPut:IMMediate
Parameters
0 | OFF | 1 | ON
*RST Value
OFF
Examples
OUTP:IMM 1
OUTP:IMM OFF
Query Syntax
OUTPut:IMMediate?
Returned Parameters
0 | 1
Related Commands
REG:STAT
REG:CURR
4.11.2
Output Protection
OUTPut:PROTection:CLEar
Available on MX/RS/BPS Series II only. This command clears the latch that disables the
output when an overvoltage (OV), overcurrent (OC), overtemperature (OT), or remote inhibit
(RI) fault condition is detected. All conditions that generated the fault must be removed
before the latch can be cleared. The output is then restored to the state it was in before the
fault condition occurred.
Command Syntax
OUTPut:PROTection:CLEar
Parameters None
Examples OUTP:PROT:CLE
Related Commands
OUTP:PROT:DEL
*RCL *SAV
4.11.3
Output Remote Inhibit Modes
OUTPut:RI[:LEVel]
This command sets the remote inhibit level mode. Factory default is LOW, which requires a
contact closure to open the output relay. The level can be reversed by setting it to HIGH.
Once set, the RI level setting is retained each time the power source is powered up. Note
that this command is only implement with firmware revision 0.28 or higher. Lower firmware
revisions only provide the default LOW setting (MX15 excluded).
Command Syntax
OUTPut:RI:LEVel
Parameters
LOW | HIGH
*RST Value
LOW
Examples OUTP:RI:LEV
HIGH
Query Syntax
OUTP:RI:LEV?
Returned Parameters
Related Commands
OUTP
NOTE: When using the Remote Inhibit input, it will be necessary to disconnect any RI
connection to the MX/RS/BPS master unit when turning on the MX/RS/BPS
master unit. During initialization, the RI connection must be OPEN or
initialization will be halted with the message WAITING FOR AUXILIARY
displayed on the LCD screen.