AMETEK BPS Series Programming Manual User Manual
Page 75

BPS / MX / RS Series SCPI Programming Manual
AMETEK Programmable Power
Manual P/N 7003-961 Rev. AA
75
OUTPut:RI:MODE
Available on MX/RS/BPS Series II only. This command selects the mode of operation of the
Remote Inhibit protection. The following modes can be selected:
LATChing
A TTL low at the RI input latches the output in the protection shutdown
state, which can only be cleared by OUTPut:PROTection:CLEar.
LIVE
The output state follows the state of the RI input. A TTL low at the RI input
turns the output off; a TTL high turns the output on.
OFF
The instrument ignores the RI input.
The RI output state is saved at power down. The factory default state is LIVE.
Command Syntax
OUTPut:RI[:MODE]
Parameters
LATChing | LIVE | OFF
*RST Value
LIVE
Examples
OUTP:RI:MODE LIVE
Query Syntax
OUTPut:RI:MODE?
Returned Parameters
Related Commands
OUTP:PROT:CLE