BPS / MX / RS Series SCPI Programming Manual

AMETEK Programmable Power

Manual P/N 7003-961 Rev. AA

75

OUTPut:RI:MODE

Available on MX/RS/BPS Series II only. This command selects the mode of operation of the
Remote Inhibit protection. The following modes can be selected:

LATChing

A TTL low at the RI input latches the output in the protection shutdown
state, which can only be cleared by OUTPut:PROTection:CLEar.

LIVE

The output state follows the state of the RI input. A TTL low at the RI input
turns the output off; a TTL high turns the output on.

OFF

The instrument ignores the RI input.

The RI output state is saved at power down. The factory default state is LIVE.

Command Syntax

OUTPut:RI[:MODE]

Parameters

LATChing | LIVE | OFF

*RST Value

LIVE

Examples

OUTP:RI:MODE LIVE

Query Syntax

OUTPut:RI:MODE?

Returned Parameters

Related Commands

OUTP:PROT:CLE