AMETEK Programmable Power

BPS / MX / RS Series SCPI Programming Manual

120

Manual P/N 7003-961 Rev AA

SYSTem:CONFigure:NOUTput

Sets the mode configuration of the power source to single-phase or a three-phase mode.
The MODE option must be enabled for this command to work. If MODE option is not
available, the query response will be FIX.

Command Syntax

SYSTem:CONFigure:NOUTput

Parameters

ONEPhase | THReephase

Example

SYST:CONF:NOUT ONE

Query Syntax

SYSTem:CONF?

Query response

ONE, THREE or FIX

Note: Query response CRD differs

from parameter format on this command.

Returned Paramters

SYSTem:ETIMe?

This command will return the total number of accumulated hours, minutes and seconds.

Command Syntax

SYSTem:ETIMe?

Parameters none
Example SYST:ETIM?
Returned Parameters

,,

SYSTem:TEMPerature?

This command will return the internal ambient temperature of the AC/DC power source in
degrees Celsius.

Command Syntax

SYSTem:TEMPerature?

Parameters none
Example SYST:TEMP?
Returned Parameters

SYSTem:COMMunicate:GPIB:ADDRess

This command will set the GPIB address.

Command Syntax

SYSTem:COMMunicate:GPIB:ADDRess

Parameters

0 to 31

Example

SYST:COMM:GPIB:ADDR 2

Returned Parameters

SYSTem:COMMunicate:LAN:ADDRess

This command will set the LAN IP address. The query format returns the IP address set. To
query the actual IP address used by the LAN interface (in case the IP address is not static
but assigned by a DHCP server), the NCON parameter (Network Connection Setting) must
be added to the query. The default value is 0.0.0.0.