AMETEK BPS Series Programming Manual User Manual
Page 108

AMETEK Programmable Power
BPS / MX / RS Series SCPI Programming Manual
108
Manual P/N 7003-961 Rev AA
VOLTage
Phase Selectable
This command programs the AC rms, DC or offset output voltage level of the power source.
The voltage mode must be set for AC to accept an AC voltage. The Voltage mode must also
be set to AC+DC to accept a DC offset value.
Command Syntax
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude][:AC]
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]:DC] +> Parameters 0 to maximum voltage range specified by the LIMit:VOLTage Unit V (rms voltage) *RST Value 0 volt Examples VOLT 250 VOLT:LEV 25 Query Syntax [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? Returned Parameters Related Commands [SOUR:]MODE VOLT:TRIG VOLTage:TRIGgered Phase Selectable This command selects the AC rms or DC amplitude that the output voltage will be set to Command Syntax SOURce:]VOLTage[:LEVel]:TRIGgered:AMPLitude] Parameters 0 to maximum voltage range specified by the LIMit:VOLTage Unit V (rms voltage) *RST Value 0 volt Examples VOLT:TRIG 120 VOLT:LEV:TRIG 120 Query Syntax SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]? Returned Parameters Related Commands VOLT VOLT:MODE
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]:OFFSet
command
during a triggered step or pulse transient.
command