AMETEK Programmable Power

BPS / MX / RS Series SCPI Programming Manual

112

Manual P/N 7003-961 Rev AA

VOLTage:SLEW:MODE

Phase Selectable

This command determines how the output voltage slew rate is controlled during a triggered
output transient. The choices are:

FIXed

The slew rate is unaffected by a triggered output transient.

STEP

The slew rate is programmed to the value set by
VOLTage:SLEW:TRIGgered when a triggered transient occurs.

PULSe

The slew rate is changed to the value set by VOLTage:SLEW:TRIGgered
for a duration determined by the pulse commands.

LIST

The slew rate is controlled by the voltage slew list when a triggered transient
occurs.

Command Syntax

[SOURce:]VOLTage:SLEW:MODE

Parameters

FIXed | STEP | PULSe | LIST

*RST Value

FIX

Examples

VOLT:SLEW:MODE LIST

VOLT:SLEW:MODE FIX

Query Syntax

[SOURce:]VOLTage:SLEW:MODE?

Returned Parameters

Related Commands

VOLT:SLEW:TRG

VOLT:SLEW

VOLTage:SLEW:TRIGgered

Phase Selectable

This command selects the slew rate that will be set during a triggered step or pulse
transient. A parameter of MAXimum will set the slew to its maximum possible rate. The
maximum allowed value is1E9.

Command Syntax

[SOURce:]VOLTage:SLEW:TRIGgered

Parameters

0 to 1E9 | MAXimum

Unit

V/S (volts per second)

*RST Value

MAXimum

Examples

VOLT:SLEW:TRIG 1

VOLT:SLEW:TRIG MAX

Query Syntax

[SOURce:]VOLTage:SLEW:TRIGgered?

Returned Parameters

Related Commands

VOLT:SLEW:MODE

VOLT:SLEW