BPS / MX / RS Series SCPI Programming Manual

64

Manual P/N 7003-961 Rev AA

MEASure:CURRent:CREStfactor?
FETCh:CURRent:CREStfactor?

Phase Selectable

These queries return the output current crest factor. This is the ratio of peak output current
to rms output current.

Query Syntax

MEASure[:SCALar]:CURRent:CREStfactor?

FETCh[:SCALar]:CURRent:CREStfactor?
Parameters None
Examples MEAS:CURR:CRES?

FETC:CURR:CRES?

Returned Parameters

Related Commands

INST:NSEL

MEASure:CURRent:HARMonic?

[3Pi Controller Only]

FETCh:CURRent:HARMonic?

[3Pi Controller Only]

Phase Selectable

These queries return the rms amplitude of the Nth harmonic of output current. The
parameter is the desired harmonic number. Queries sent with a value of 0 return the dc
component. A value of 1 returns the fundamental output frequency. Harmonic frequencies
can be queried up to the fundamental measurement bandwidth of the measurement system:

MX Series I:

25.6 microseconds for single-phase mode and 76.8 microseconds for three-

phase mode.

MX/RS Series II: 10.4 microseconds for single-phase mode and 31.2 microseconds for
three-phase mode.

Thus the maximum harmonic that can be measured is dependent on the output frequency.
Any harmonics that represent frequencies greater than above frequencies are returned as 0.

Query Syntax

MEASure[:SCALar]:CURRent:HARMonic[:AMPLitude]?

FETCh[:SCALar]:CURRent:HARMonic[:AMPLitude]?
Parameters

0 to 50

Examples MEAS:CURR:HARM?

3

FETC:CURR:HARM?

1

Returned Parameters

Related Commands

INST:NSEL