BPS / MX / RS Series SCPI Programming Manual

AMETEK Programmable Power

Manual P/N 7003-961 Rev. AA

49

4.2.8

Output - DC Voltage

CALibrate[:SOURce]:VOLTage:DC:LRANge[:FSCale]

This command will result in the calibration coefficent being calculated for the DC full-scale
output voltage in the low voltage range.

Command Syntax

CALibrate:VOLTage:DC:LRANge

Parameters

(programmed output voltage)

Examples CAL:VOLT:DC:LRAN

-2

Query Syntax

CALibrate:VOLTage:DC:LRANge?

Returned Parameters

(cal coefficient value between 0.9 and 1.1)

Related Commands

CAL:VOLT:DC:LRAN:NEG

CALibrate[:SOURce]:VOLTage:DC:LRANge:ZERO

This command will set the calibration coefficent for the dc output voltage offset at the low
voltage range.

Command Syntax

CALibrate:VOLTage:DC:LRANge:ZERO

Parameters

(a value between -127 and +128)

Examples CAL:VOLT:DC:LRAN:ZERO

+10

Query Syntax

CALibrate:VOLTage:DC:LRANge:ZERO?

Returned Parameters

(value range -127 to +128)

Related Commands

CAL:VOLT:DC:LRAN:ZERO

CALibrate[:SOURce]:VOLTage:DC:HRANge[:FSCale]

This command will result in the calibration coefficent being calculated for the DC full scale
output voltage in the high voltage range.

Command Syntax

CALibrate:VOLTage:DC:HRANge

Parameters

(programmed output voltage)

Examples CAL:VOLT:DC:HRAN

-2

Query Syntax

CALibrate:VOLTage:DCHRANge?

Returned Parameters

(cal coefficient value between 0.9 and 1.1)

Related Commands

CAL:VOLT:DC:LRAN:ZERO

CALibrate[:SOURce]:VOLTage:DC:HRANge:ZERO

This command will set the calibration coefficent for the dc output voltage offset at the high
voltage range.

Command Syntax

CALibrate:VOLTage:DC:HRANge:ZERO

Parameters

(a value between -127 and +128)

Examples CAL:VOLT:DC:HRAN:ZERO +10
Query Syntax

CALibrate:VOLTage:DC:HRANge:ZERO?

Returned Parameters

(value range -127 to +128)

Related Commands

CAL:VOLT:DC:LRAN