AMETEK BPS Series Programming Manual User Manual
Page 161

BPS / MX / RS Series SCPI Programming Manual
AMETEK Programmable Power
Manual P/N 7003-961 Rev. AA
161
6.6.10
Selecting the Measurement Trigger Source
The trigger system is waiting for a trigger signal in the Initiated state. Before you generate a
trigger, you must select a trigger source. To select the SYNC Trigger to start the
measurement at specific phase angles in the waveform:
TRIGger:ACQuire:SOURce SYNC
TRIGger:SYNC:SOUR PHAS
TRIGger:SYNC:PHASE 0
To select IEEE-488 bus triggers (group execute trigger, device trigger, or *TRG command),
use:
TRIGger:ACQuire:SOURce BUS
To select the signal driving the Trigger Out from the transient, use:
TRIGger:ACQuire:SOURce TTLTrg
6.6.11
Generating Measurement Triggers
Providing that you have specified the appropriate trigger source, you can generate triggers
as follows:
• By sending one of the following over the IEEE-488:
• TRIGger:ACQuire:IMMediate
• *TRG
• a group execute trigger (GET)
• By waiting for a specific phase angle of the waveform
• By generating an output transient that causes the Trig Out to output a pulse.
• By pressing the front panel START soft key in the HARMONICS/WAVEFORM when the
unit is operating in local mode.