AMETEK BPS Series Programming Manual User Manual
Page 50

AMETEK Programmable Power
BPS / MX / RS Series SCPI Programming Manual
50
Manual P/N 7003-961 Rev AA
4.2.9
Output - Impedance
[MX15-1Pi/MX30-3Pi/MX45-3Pi/RS only]
CALibrate[:SOURce]:REAL[:FSCale]
This command will set the calibration coefficent for the real part of the output impedance.
Command Syntax
CALibrate[:SOURce]:REAL[:FSCale]
Parameters
Examples
CAL:REAL +10
Query Syntax
CALibrate[:SOURce]:REAL[:FSCale]?
Returned Parameters
Related Commands
CALibrate[:SOURce]:REACtive[:FSCale]
CALibrate[:SOURce]:REAL:ZERO
This command will set the lowest real part of the output impedance that could be
programmed.
Command Syntax
CALibrate[:SOURce]:REAL:ZERO
Parameters
Examples
CAL:REAL:ZERO 100
Query Syntax
CALibrate[:SOURce]:REAL:ZERO?
Returned Parameters
Related Commands
CALibrate[:SOURce]:REACtive:ZERO
CALibrate[:SOURce]:REACtive[:FSCale]
This command will set the calibration coefficent for the reactive part of the output
impedance.
Command Syntax
CALibrate[:SOURce]:REACtive[:FSCale]
Parameters
Examples
CAL:REAL +10
Query Syntax
CALibrate[:SOURce]:REACtive[:FSCale]?
Returned Parameters
Related Commands
CALibrate[:SOURce]:REAL[:FSCale]
CALibrate[:SOURce]:REACtive:ZERO
This command will set the lowest reactive part of the output impedance that could be
programmed.
Command Syntax
CALibrate[:SOURce]:REACtive:ZERO
Parameters
Examples
CAL:REACtive:ZERO 100
Query Syntax
CALibrate[:SOURce]:REACtive:ZERO?
Returned Parameters
Related Commands
CALibrate[:SOURce]:REAL:ZERO