AMETEK Programmable Power

BPS / MX / RS Series SCPI Programming Manual

48

Manual P/N 7003-961 Rev AA

CALibrate[:SOURce]:VOLTage[:AC]:HRANge:ZERO

This command will set the calibration coefficent for the output voltage offset at the high
voltage range.

Command Syntax

CALibrate:VOLTage:HRANge:ZERO

Parameters

(a value between -127 and +128)

Examples

CAL:VOLT:HRAN:ZERO +10

Query Syntax

CALibrate:VOLTage:HRANge:ZERO?

Returned Parameters

(value range -127 to +128)

Related Commands

CAL:VOLT:HRAN CAL:VOLT:HRAN:HFR

CALibrate[:SOURce]:VOLTage[:AC]:HRANge:HFRequency

This command will set the calibration coefficent for the AC full scale output voltage at the
high voltage range and at high output frequency

Command Syntax

CALibrate:VOLTage:HRANge:HFRequency

Parameters

(a value between -127 and +128)

Examples CAL:VOLT:HRAN:HFRequency

+10

Query Syntax

CALibrate:VOLTage:HRANge:HFRequency?

Returned Parameters

(value range -127 to +128)

Related Commands

CAL:VOLT:HRAN:ZERO CAL:VOLT:HRAN

CALibrate[:SOURce]:VOLTage:HFRrequency

This query command retrieves the harmonic calibration coefficients for the IEC413 option.
Available only on MX/RS/BPS Series II models with firmware revision 4.20 or higher. This
query returns a comma-separated list of seven calibration coefficients. These coefficients
can only be set by performing a voltage measurement calibration.

Query Syntax

CALibrate:VOLTage:HFRequency?

Returned Parameters

Related Commands

CAL:IHAR?

CALibrate[:SOURce]:IHARmonic

This query command retrieves the interharmonic calibration coefficient for the IEC413
option. This query returns a comma-separated list of the calibration coefficient and the
frequency at which the calibration was peformed.

Query Syntax

CALibrate:IHARmonic?

Returned Parameters

Related Commands

CAL:VOLT:HFR