BPS / MX / RS Series SCPI Programming Manual
AMETEK Programmable Power
Manual P/N 7003-961 Rev. AA
63
MEASure:CURRent:DC?
FETCh:CURRent:DC?
Phase Selectable
These queries return the dc component of the output current being sourced at the output
terminals. This command should be used when the voltage mode of the source is set for DC
Query Syntax
MEASure[:SCALar]:CURRent:DC?
FETCh[:SCALar]:CURRent:DC?
Parameters None
Examples MEAS:CURR?
FETC:CURR?
Returned Parameters
Related Commands
INST:NSEL
MEASure:CURRent:AMPLitude:MAXimum?
FETCh:CURRent:AMPLitude:MAXimum?
Phase Selectable
These queries return and hold the absolute value of the peak current as sampled over one
measurement acquisition of 4096 data points. The returned value will be updated only when
a larger value is found. To update the value with every measurement a peak current reset
command should be used prior to the peak measurements.
Query Syntax
MEASure[:SCALar]:CURRent:AMPLitude:MAXimum?
FETCh[:SCALar]:CURRent:AMPLitude:MAXimum?
Parameters None
Examples MEAS:CURR:AMPL:MAX?
FETC:CURR:AMPL:MAX?
Returned Parameters
Related Commands
INST:NSEL
MEAS:CURR:AMPL:RESet
MEASure:CURRent:AMPLitude:RESet
Phase Selectable
This command will reset the peak current measurement to zero.
Query Syntax
MEASure[:SCALar]:CURRent:AMPLitude:RESset.
Parameters None
Examples MEAS:CURR:AMPL:RES
Returned Parameters
None
Related Commands
MEAS:CURR:AMPL:MAX?