IEC413:HARMonics:TABle
This command allows selection of only one of the Individual harmonics tables of the IEC
61000-4-13 standard. Default is to run all tables but some EUT’s for class 3 may require
only Table 1 to be run. Available selections are Table 1, 2, 3 or all tables.
Command Syntax
IEC413:HARMonics:TABle
Parameters
1 | 2 | 3 | ALL
Examples IEC413:HARM:TAB
ALL
Query Syntax
IEC413:HARM:TAB?
Returned Parameters
IEC413:HARMonics:DWELl
This command will set the dwell time in seconds for each harmonic. The class must be
selected prior to this command. Query with min and max to find the range of the dwell time.
Command Syntax
IEC413:HARM:DWELl
Parameters
Examples IEC413:HARM:DWEL
10
Query Syntax
IEC413:HARM:DWEL?
Returned Parameters
IEC413:HARMonics:PAUSe
This command will set the pause time in seconds. This is the time between each harmonic.
The class must be selected prior to this command. Query with min and max to find the
range of the dwell time.
Command Syntax
IEC413:HARM:PAUSe
Parameters
Examples IEC413:HARM:PAUS
2
Query Syntax
IEC413:HARM:PAUS?
Returned Parameters
IEC413:IHARMonics:LEVel
This command will set the level for the interharmonics test. The command will accept either
one parameter or five parameters. In case of one parameter, the level will be set for the
selected range only. The IEC413:STATe must be in the off position and the user class must
be selected to accept the command.
Command Syntax
IEC413:IHARmonics:LEVel
Parameters
0 to 20%
Examples IEC413:IHAR:LEV
10
Query Syntax
IEC413:IHAR:LEV?
Returned Parameters
IEC413:IHARmonics:DWELl
This command will set the dwell time in seconds for each interharmonics frequency. The
class must be selected prior to this command. Query with min and max to find the range of
the dwell time.