BPS / MX / RS Series SCPI Programming Manual

204

Manual P/N 7003-961 Rev AA

REGenerate:OUTPut[:STATe]

This command enables or disables the source regenerate mode. When the regenerate
mode is OFF (disabled), the MX/RS behaves like a standard MX/RS without the SNK option.
In order to turn on the REGENERATE mode of operation (enable), the output relay has to be
open and the programmed frequency has to be within a 40 – 80Hz range. The query form
returns the regenerate state.

Command Syntax

REGenerate:OUTPut[:STATe]

Parameters

0 | OFF | 1 | ON

Examples

REG:OUTP 1 REG:OUTP:STAT ON

Query Syntax

REG:OUTPut[:STATe]?

Returned Parameters

0 | 1

REGenerate:OVER[:VOLTage]

This command will set or query the upper voltage limit over which the EUT is allowed to
operate.

Command Syntax

REGenerate:OVER[:VOLTage]

Parameters

0 to max. value for selected voltage range.

Examples REG:OVER:VOLT

120

Query Syntax

REGenerate:OVER[:VOLTage]?

Returned Parameters

REGenerate:UNDer[:VOLTage]

This command will set or query the lower voltage limit over which the EUT is allowed to
operate.

Command Syntax

REGenerate:UNDer[:VOLTage]

Parameters

0 to max. value for selected voltage range.

Examples REG:UND:VOLT

120

Query Syntax

REGenerate:UNDer[:VOLTage]?

Returned Parameters