Manual P/N 7003-961 Rev AA
TRIGger[:TRANSient]:SOURce
This command will determine the source of the trigger event for the transient. The following
sources are available:
• IMMediate Triggering occurs as soon as the INIT command is received.
• BUS
Triggering occurs following the INIT command after receiving the *TRG
command or a Group Execute Trigger (GET) IEEE signal.
• EXTernal Triggering occurs when an external trigger input is received.
Command Syntax
TRIGger[:TRANSient]:SOURce
Parameters
IMMediate | BUS | EXT
*RST Value
IMMediate
Examples TRIG:SOUR
BUS
Related Commands
ABOR
TRIGger:COUNt
This command determines whether the phase synchronization is accomplished at the
begining of the transient list or pulse for every COUNt or at the begining of the transient for
the first count only. It accepts the following parameters.
• NONe
Synchronization is accomplished once
• ALL
Synchronization is accomplished for every count.
Command Syntax
TRIGger:COUNt ALL
TRIGger:COUNt
NONE
Parameters ALL,
NONE
*RST Value
NONE
Examples TRIG:COUN
ALL
TRIG:COUN
NONE
Query Syntax
TRIG:COUN?
Returned Parameters
Related Commands
ABOR TRIG
TRIG:DEL
TRIG:SYNC
TRIG:SYNC:PHAS
INIT
INIT:CONT *TRG WAI